Higher accuracy with fewer projections? Automated scan angle selection for dimensional Computed Tomography based on a simple data completeness measure for the part surface
L. Butzhammer, M.R.O. Braun, C. Herath, T. Hausotte, in: E-Journal of Nondestructive Testing, NDT.net GmbH & Co. KG, 2026.
DOI
Conference Paper
| Published
| English
Author
Butzhammer, Lorenz;
Braun, Matthias Robert Oskar;
Herath, Colin;
Hausotte, Tino
Department
Project
Abstract
<jats:p>Selecting scan angles such that surface segments are aligned with straight X-ray paths (i.e., rays are tangential to the surface and therefore perpendicular to the local surface normal) is known to produce sharper transitions of those surface segments in the reconstructed volume. This enhances dimensional accuracy in sparse-view computed tomography (CT). However, existing approaches offer no direct means to exploit this criterion for automatic scan-angle optimization. We propose a method that uses a virtual representation of the CT setup, including an STL surface model of the inspected part, to automatically identify taskspecific scan angles. Using elementary vector calculus, the algorithm determines projection directions that generate tangential X-rays for targeted surface segments. To support different levels of geometric complexity, we introduce two variants of the angle-selection procedure. The methods were experimentally validated on two objects with distinct absorption and geometric characteristics. For a steel gauge block, employing the minimum number of task-specific projections required for surface-data completeness substantially outperformed a conventional high-projection scan. For a geometrically more complex test object, surface-related errors were still reduced within the region of interest. The proposed approach – particularly suited for flat surface structures and not accounting for image-degrading factors other than cone-beam artifacts – shows promise for high-throughput dimensional metrology of mono-material parts.</jats:p>
Publishing Year
Proceedings Title
e-Journal of Nondestructive Testing
Volume
31
Issue
3
Conference
15th Conference on Industrial Computed Tomography (iCT)
Conference Location
Linz
ISSN
LibreCat-ID
Cite this
Butzhammer L, Braun MRO, Herath C, Hausotte T. Higher accuracy with fewer projections? Automated scan angle selection for dimensional Computed Tomography based on a simple data completeness measure for the part surface. In: E-Journal of Nondestructive Testing. Vol 31. NDT.net GmbH & Co. KG; 2026. doi:10.58286/32560
Butzhammer, L., Braun, M. R. O., Herath, C., & Hausotte, T. (2026). Higher accuracy with fewer projections? Automated scan angle selection for dimensional Computed Tomography based on a simple data completeness measure for the part surface. E-Journal of Nondestructive Testing, 31(3). https://doi.org/10.58286/32560
@inproceedings{Butzhammer_Braun_Herath_Hausotte_2026, title={Higher accuracy with fewer projections? Automated scan angle selection for dimensional Computed Tomography based on a simple data completeness measure for the part surface}, volume={31}, DOI={10.58286/32560}, number={3}, booktitle={e-Journal of Nondestructive Testing}, publisher={NDT.net GmbH & Co. KG}, author={Butzhammer, Lorenz and Braun, Matthias Robert Oskar and Herath, Colin and Hausotte, Tino}, year={2026} }
Butzhammer, Lorenz, Matthias Robert Oskar Braun, Colin Herath, and Tino Hausotte. “Higher Accuracy with Fewer Projections? Automated Scan Angle Selection for Dimensional Computed Tomography Based on a Simple Data Completeness Measure for the Part Surface.” In E-Journal of Nondestructive Testing, Vol. 31. NDT.net GmbH & Co. KG, 2026. https://doi.org/10.58286/32560.
L. Butzhammer, M. R. O. Braun, C. Herath, and T. Hausotte, “Higher accuracy with fewer projections? Automated scan angle selection for dimensional Computed Tomography based on a simple data completeness measure for the part surface,” in e-Journal of Nondestructive Testing, Linz, 2026, vol. 31, no. 3, doi: 10.58286/32560.
Butzhammer, Lorenz, et al. “Higher Accuracy with Fewer Projections? Automated Scan Angle Selection for Dimensional Computed Tomography Based on a Simple Data Completeness Measure for the Part Surface.” E-Journal of Nondestructive Testing, vol. 31, no. 3, NDT.net GmbH & Co. KG, 2026, doi:10.58286/32560.
All files available under the following license(s):
Copyright Statement:
This Item is protected by copyright and/or related rights. [...]
Link(s) to Main File(s)
Access Level
Closed Access